Page 23 - HIGHWAY ENGINEERING LABORATORY_FLIP BUILDER
P. 23

4.0  DETAILED PROCEDURE :  POLISHED STONE VALUE (PSV)

                                                                                       REFERENCE
                                      WORK PROCESS
                                                                                      DOCUMENT
                 the scale. Ensure that the track of the slider is parallel to the

                 long edge of the specimen across the sliding distance.
               • Use the following procedure to ensure that the rubber slider is
                 correctly conditioned.
               • NOTE It is possible to reverse the slider to make use of both
                 edges.

                 a) Maintain a stock of Criggion specimens for calibration
                    purposes. These shall be made from stone from Criggion
                    Quarry and shall be made and polished as in an actual

                    determination. When tested they shall yield a value in the
                    range58 to66. Record the values and air-dry the specimens
                    and store them in a sealed container for future use.
                 b) Before using a new slider, swing it five times over the dry

                    surface of a polished Criggion specimen following with a
                    further 20 swings on its wetted surface. Keep the Criggion
                    specimen used for this purpose apart from the Criggion
                    calibration specimens tested in it may be used repeatedly

                    provided its value (when wet) does not fall below55.
                 c) Before measuring the set of specimens polished on each test
                    run, check the performance of the slider by testing a
                    Criggion calibration specimen. Record the resulting value.

                    Additional  polishing  through   repeated    testing  yields
                    progressively lower values and the control shall be discarded
                    when its value falls below 57. A new control shall then be
                    drawn from the stock. If a check value is more than 2 units

                    lower (or one unit higher) than the last recorded value for
                    the control, discontinue testing and rectify any fault in the
                    instrument or its operation or due to changes in the slider. It is
                    recommended that more than one slider be kept in use to

                    help differentiate between a faulty slider and a defective
                    instrument.
                 d) Discard any slider that develops excessive burring and
                    scoring through prolonged use.




              Higway Engineering Laboratory                                                                      page      10
   18   19   20   21   22   23   24   25   26   27   28